Assessment of stroke volume and cardiac output by Doppler echocardiography

P Niederle1, A Stárek, V Jezek

  • 1Institute of Physiological Regulations, Czechoslovak Academy of Sciences, Prague.

Cor Et Vasa
|January 1, 1988
PubMed

Insights

Doppler ultrasound accurately measures stroke volume and cardiac output, correlating closely with traditional hemodynamic methods. This non-invasive technique is a reliable alternative for assessing cardiac function in patients with good echo Doppler recordings.

Area of Science:

  • Cardiovascular Physiology
  • Medical Imaging
  • Diagnostic Ultrasound

Background:

  • Accurate measurement of stroke volume and cardiac output is crucial for assessing cardiovascular function.
  • Traditional methods like Fick's principle and thermodilution are invasive or have limitations.
  • Non-invasive techniques are sought for routine clinical assessment.

Purpose of the Study:

  • To evaluate the accuracy and reliability of pulse and continuous-wave Doppler ultrasound for measuring stroke volume and cardiac output.
  • To compare Doppler-derived hemodynamic values with those obtained using Fick's principle of thermodilution.

Main Methods:

  • Utilized pulse and continuous-wave Doppler technique for 51 measurements in 37 individuals.
  • Compared Doppler results with simultaneous or immediately subsequent measurements using Fick's principle of thermodilution.
  • Performed regression analysis and paired t-tests to assess correlation and agreement.

Main Results:

  • Demonstrated a close correlation between Doppler and hemodynamic values (r=0.86 for stroke volume, r=0.94 for cardiac output; p<0.001).
  • Paired t-test showed no significant difference between Doppler and reference methods.
  • Standard deviation of Doppler stroke volume was 10%, with only two measurements differing by over 25%.

Conclusions:

  • Doppler ultrasound is a highly accurate and reliable non-invasive method for determining stroke volume and cardiac output.
  • It shows excellent agreement with established hemodynamic assessment techniques.
  • Recommended as an alternative method for cardiac output determination in patients with high-quality echo Doppler recordings.