Related Experiment Video
Updated: Dec 24, 2025

Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Assessment of low plasma concentrations of apixaban in the periprocedural setting
Sarah Lessire1, Anne-Sophie Dincq1, Romain Siriez2
1Université catholique de Louvain, CHU UCL Namur, Namur Thrombosis and Hemostasis Center (NTHC), NAmur Research Institute for LIfe Sciences (NARILIS), Department of Anesthesiology, Yvoir, Belgium.
Introduction:
Estimation of residual apixaban plasma concentrations may be requested in the management of emergencies. This study aims at assessing the performance of specific anti-Xa assays calibrated with apixaban on real-life samples with low apixaban plasma concentrations (<30 ng/mL) and on-treatment ranges, with and without interference of low-molecular-weight heparin (LMWH).
Methods:
The performance of the STA® -Liquid Anti-Xa assay (STA® LAX) and the low and normal procedures of the Biophen® Direct Factor Xa Inhibitors (DiXaI) assay was tested on 134 blood samples, collected from patients on apixaban, wherefrom 74 patients received LMWH after apixaban cessation. The results were compared with the liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) measurements.
Results:
The Biophen® DiXaI, Biophen® DiXaI LOW, and STA® LAX showed very good correlation with LC-MS/MS measurements in patients without LMWH administration (Spearman r .95, .99, and .98, respectively). Their limits of quantitation were defined at 48, 24, and 12 ng/mL, respectively. The Bland-Altman test measured mean bias (SD) at 5.6 (13.1), -2.5 (5.0), and -0.8 (6.1) ng/ml, respectively. The Spearman r of the Biophen® DiXaI decreased to 0.64 in presence of low apixaban concentrations. The Spearman r of the Biophen® DiXaI LOW and STA® LAX decreased to 0.39 and 0.26, respectively, in presence of LMWH.
Conclusions:
The accuracy of the low methodologies (Biophen® DiXaI LOW and STA® LAX) is slightly improved for low apixaban plasma concentrations, compared with the normal procedure of Biophen® DiXaI. The interference of LMWH on the low methodologies is measurable, however, less important than the previously reported interference of LMWH on rivaroxaban calibrated specific anti-Xa assays.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Venous Thrombosis III: Interprofessional Care
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Drug Concentrations: Measurements
Plasma...
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care