Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Improved scan method for dental imaging using multidetector computed tomography: a phantom study
Yuki Sakai1, Kazutoshi Okamura2, Erina Kitamoto2
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan.
Objectives:
This study aimed to propose an improved scan method to shorten irradiation time and reduce radiation exposure.
Methods:
The maxilla of a human head CT phantom and a Catphan phantom were used for qualitative and quantitative assessment, respectively. The phantoms were scanned by a 160-row multidetector CT scanner using volumetric and helical scanning. In volumetric scanning, the tube current varied from 120 to 60 to 30 to 20 mA with a tube voltage of 120 kV. Images were reconstructed with a bone kernel using iterative reconstruction (IR) and filtered back projection. As a reference protocol, helical scanning was performed using our clinical setting with 120 kV. Two dental radiologists independently graded the quality of dental images using a 4-point scale (4, superior to reference; 1, unacceptable). For the quantitative assessment, we assessed the system performance from each scan.
Results:
There was no significant difference between the image quality of volumetric scanning using the 60 mA protocol reconstructed with IR and that of the reference (3.08 and 3.00, p = 0.3388). The system performance values at 1.0 cycles/mm of volumetric scanning and 60 mA protocol reconstructed with IR and reference were 0.0038 and 0.0041, respectively. The effective dose of volumetric scanning using the 60 mA protocol was 51.8 µSv, which is a 64.2% reduction to that of the reference.
Conclusions:
We proposed an improved scan method resulting in a 64.2% reduction of radiation dose with one-fourth of irradiation time by combining volumetric scanning and IR technique in multidetector CT.
More Related Videos
09:10Digital Hybrid Model Preparation for Virtual Planning of Reconstructive Dentoalveolar Surgical Procedures
Published on: August 5, 2021
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...