Comparative study of the Orthofix and Pittsburgh frames for external fixation of unstable pelvic ring fractures
A L Bell1, R A Smith, T D Brown
1Department of Orthopaedic Surgery, University of Iowa Hospitals and Clinics, Iowa City 52242.
Abstract:
We compared the mechanical performance of the Orthofix pelvic external fixator with that of the Pittsburgh triangular frame (PTF) on eight fresh-frozen cadaver pelves with experimentally created Malgaigne (double-vertical) fracture/dislocations. The pelves were quasi-statically loaded in longitudinal compression and loss of reduction (i.e., failure) was defined as 1.5 cm of diastasis at either of the fracture/dislocation sites. The Orthofix frame-pelvis complex (four 6-mm half-pins) was comparable to the PTF-pelvis complex (eight 5-mm half-pins) in terms of load to failure and overall stiffness, but was 4-7 times stiffer at lower load levels. Both frame-and-pin complexes deformed moderately at the pin-bone interface, but, while the PTF also showed moderate displacement between frame components, the Orthofix had almost no frame intercomponent motion.