Related Experiment Video
Updated: Sep 6, 2026

The α-test: Rapid Cell-free CD4 Enumeration Using Whole Saliva
Published on: May 16, 2012
Evaluation of the automated aptima HIV-1 quant Dx assay for both qualitative and quantitative testing in pediatric
Priyanka Uprety1, Darcy Curtis2, Ana María Cárdenas1
1Infectious Disease Diagnostics Laboratory, Children's Hospital of Philadelphia, Philadelphia, PA, USA; Department of Pathology and Laboratory Medicine, Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA.
Background:
There are currently no FDA-cleared assays with a dual-claim for both diagnosis and monitoring of HIV-1. The Aptima HIV-1 Quant Dx Assay on the Panther platform (Panther) is the first commercially available test that is CE-marked for both HIV-1 diagnosis and monitoring, but only FDA-cleared for HIV-1 monitoring.
Objective:
To evaluate the Panther assay for use as a qualitative and quantitative HIV-1 assay in a pediatric population, including patients younger than 24 months old, and review its effect on laboratory efficiency and hands-on-time following its implementation.
Study Design:
100 patient specimens previously tested on the Abbott m2000 RealTime HIV-1 assay (RealTime) and 185 patient specimens previously tested on the Aptima HIV-1 RNA Qualitative Assay (RNA Qual) were tested on the Panther. Verification panels were used to establish precision and linearity. In addition, 268 samples from 134 patients under 24 months of age were also evaluated on the Panther.
Results:
Overall agreement between the Panther and RealTime assays was 83 %. The mean difference between the two methods was 0.10 Log copies/mL. All Panther measurements were linear across the dynamic range (R2 = 0.999). The Panther and RNA Qual assays showed 100 % agreement. Implementation of the assay opened 600 sq. ft. of space, saved 0.4 FTE and reduced hands-on-time by 70 %.
Conclusions:
The Panther assay is an excellent option for HIV-1 qualitative detection and quantitative testing in pediatric patients, including those under 24 months of age. HIV testing on one platform has opened up space in the clinical laboratory and reduced hands on testing time.