[Organ-based Tube-current Modulation Applied on Different MDCT Scanners: Reduction in the Radiation Dose to the Eye Lens at Head CT]
Related Experiment Video
Updated: Dec 23, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Nobuo Kitera1,2, Kosuke Matsubara3, Chikako Fujioka1
1Department of Radiology, Hiroshima University Hospital.
Purpose:
Organ-based tube current modulation (OB-TCM) techniques, which are provided by three vendors, reduces the radiation dose to the lens of the eyes by decreasing the tube current, when the X-ray tube passes over the anterior surface of critical organs. However, the characteristics of dose modulation of these techniques are different. The purpose of this study was to understand the performance characteristics of OB-TCM technique of each computed tomography (CT) vendor at head CT.
Methods:
We used three CT scanners (SOMATOM Definition Flash; Siemens Healthcare, Revolution CT; GE Healthcare, and Aquilion ONE Genesis Edition; Canon Medical Systems). We measured the radiation dose to the lens surface as evaluation of radiation dose reduction and measured the image noise as index of image quality. We measured the radiation dose rate in the air for analysis of the characteristics of dose modulation in each OB-TCM.
Results:
When applying OB-TCM, the radiation doses for the lens surface were decreased by 28%, 22%, and 25% for Siemens, GE, and Canon CT scanners, respectively, and the image noise level was increased by 5.6%, 8.5%, and 15.1% for Siemens, GE, and Canon CT scanners, respectively. The characteristics of dose modulation in each OB-TCM were also confirmed by measured the radiation dose rate.
Conclusion:
We confirmed that each OB-TCM has different influence on image quality and radiation doses for lens surface, due to the different characteristics of dose modulation for each CT vendor.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...