Performance of factor IX extended half-life product measurements in external quality control assessment programs
Angelique Nederlof1,2, Steve Kitchen3, Piet Meijer4
1Department of Haematology, Erasmus University Medical Centre Rotterdam, Rotterdam, the Netherlands.
Background:
Patients with hemophilia B are increasingly treated with extended half-life (EHL) factor IX (FIX) concentrates. For the laboratory, introduction of these EHL concentrates presents a major challenge. To understand the variation in FIX activity levels, all available diagnostic assays need to be directly compared.
Methods:
The ECAT, UKNEQAS, and RCPAQAP have collaboratively performed a global survey to evaluate the quality of FIX measurements using FIX deficient plasma samples spiked with recombinant FIX (rFIX), rFIXFP, rFIXFc, and N9-GP to levels at typical FIX trough (6 IU/dL) and peak levels (60 IU/dL). Participants were asked to use their routine protocols, using one-stage assays (OSA) or chromogenic assays (CA).
Results:
In samples spiked with 6 IU/dL product, median (25%-75% range) FIX activity levels (OSA), were 8.0 IU/dL (7.0-9.2) for rFIX, 6.0 IU/dL (4.0-7.1) for rFIXFP, 6.6 IU/dL (5.5-8.0) for rFIXFc, and 4.9 IU/dL (3.5-8.4) for N9-GP. In samples spiked with 60 IU/dL, FIX activity levels measured (using OSA) was 63.0 IU/dL (59.9-67.0) for rFIX, 42.5 IU/dL (28.2-47.0) for rFIXFP, 50.0 IU/dL (45.0-55.0) for rFIXFc, and 34.0 IU/dL (24.8-67.5) for N9-GP. Considerable differences were observed between reagents for all samples. With CA, there was also quite some variation, but no differences between reagents.
Conclusion:
Large variation is observed in the measurement of FIX activity levels after administration of rFIX and EHL FIX products. For N9-GP, most silica-based assays show especially high levels. It is essential to standardize and improve reliability of measurements of these concentrates as diagnosis and treatment monitoring is based on these results.