Early malfunction of a biliary self-expandable metal stent with an antireflux valve: A case report
Related Experiment Video
Updated: Dec 23, 2025

Endoscopic Ultrasound-Guided Biliary Drainage: Endoscopic Ultrasound-Guided Hepaticogastrostomy in Malignant Biliary Obstruction
Published on: March 25, 2022
Sang Hoon Kim1, Chi Hyuk Oh2, Jae Min Lee1
1Division of Gastroenterology and Hepatology, Department of Internal Medicine, Korea University College of Medicine.
Rationale:
Biliary drainage is essential to resolve jaundice in patients with malignant biliary obstruction. Recently, a biliary self-expandable metal stent (SEMS) with an antireflux valve was developed to prevent enteric-biliary reflux. Its antireflux valve was made of expanded polytetrafluoroethylene (ePTFE), a biostable and biocompatible material. Changes of the ePTFE membrane of medical devices are rarely reported in clinical practice.
Patient Concerns:
A 59-year-old woman was admitted with a complaint of jaundice. Around 1 month before, she underwent endoscopic biliary stenting using a SEMS with an antireflux valve.
Diagnosis:
Acute cholangitis due to stent clogging was detected on abdominal computed tomography.
Interventions:
Endoscopic retrograde cholangiopancreatography and removal of the SEMS with deformity were performed.
Outcomes:
The obstructive jaundice was rapidly resolved after the removal of the clogged stent.
Lessons:
Although the ePTFE membrane is widely used to make medical devices, unexpected changes could occur after the membrane is exposed to bile juice in the duodenum. When performing biliary stenting using a SEMS with an antireflux valve, both the risk of early malfunction and the benefit of the antireflux effect should be considered.