Automatic Derivation of Continuous Positive Airway Pressure Settings: Comparison with In-Laboratory Titration
Hyeyun Kim1, Mija Lee2, Young Hwangbo3
1Department of Neurology, Catholic Kwandong University College of Medicine, International St. Mary's Hospital, Incheon, Korea.
Background And Purpose:
This study was designed to investigate differences in the final recommended pressure setting between that derived from an autotitrating continuous positive airway pressure (APAP) device and manual in-laboratory continuous positive airway pressure (CPAP) titration, as well as the factors that influence pressure differences in patients with obstructive sleep apnea (OSA).
Methods:
This study enrolled 50 patients with OSA. All patients underwent both APAP titration and manual CPAP titration. We obtained the average device pressure ≤90% of the time (APAP₉₀) from the downloaded manual for the APAP machine and the optimal pressure obtained by manual CPAP titration (CPAPmanual). We placed the subjects into three groups based on the pressure difference (ΔP) obtained by subtracting CPAPmanual from APAP₉₀: 1) Prequal (ΔP=0), 2) CPAPmanual+ (ΔP ≤-1), and 3) APAP₉₀+ (ΔP ≥1). Regression analysis was conducted to identify predictive factors associated with ΔP.
Results:
The values of APAP₉₀ and CPAPmanual were 9.50±3.03 cmH₂O and 9.48±2.71 cmH₂O (mean±SD), respectively (p=0.95). The Prequal, CPAPmanual+, and APAP₉₀+ groups comprised 9 (18%), 23 (46%), and 18 (36%) subjects, respectively. Regression analyses revealed that male sex [β=3.539, 95% confidence interval (CI)=0.040-7.039], body mass index (BMI) (β=0.186, 95% CI=0.020-0.352), and average usage per day (β=0.768, 95% CI=0.077-1.459) were associated with ΔP.
Conclusions:
While the mean pressure in the overall cohort did not differ significantly between APAP₉₀ and CPAPmanual, there was a discordance majority showing different single pressures obtained when applying the two titration methods. Being Male, having an increased BMI, and having an increased average usage per day of APAP were significantly correlated with increased ΔP in this study.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...