Related Experiment Video
Updated: Dec 23, 2025

07:42
Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
648
Dentinal Microcrack Formation after Root Canal Instrumentation by XP-Endo Shaper and ProTaper Universal: A
Sarah M Alkahtany1, Ebtissam M Al-Madi1
1Division of Endodontics, Department of Restorative Dentistry, College of Dentistry, King Saud University, P.O. Box 5967, Riyadh 11432, Saudi Arabia.
International Journal of Dentistry
|April 24, 2020
Summary
This study compared XP-endo shaper (XPES) and ProTaper Universal (PTU) for root canal instrumentation. While both systems showed no significant difference in dentinal microcracks in certain root thirds, XPES created more incomplete microcracks in the middle third.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Root canal instrumentation is crucial for endodontic treatment success.
- Different rotary file systems may influence dentinal tissue integrity.
- Understanding microcrack formation is vital for preventing root fractures.
Purpose of the Study:
- To compare dentinal microcrack formation after root canal instrumentation using XP-endo shaper (XPES) versus ProTaper Universal (PTU).
- To evaluate the impact of these instruments on dentinal integrity at body temperature.
- To analyze microcrack formation in different root thirds using microcomputed tomography.
Main Methods:
- Nineteen mandibular molars with Vertucci Type IV canals were instrumented with either XPES or PTU.
- Microcomputed tomography (micro-CT) was used to scan roots before and after instrumentation.
- Dentinal microcracks were quantified in cervical, middle, and apical thirds.
Main Results:
- Both XPES and PTU increased microcracks in the cervical third.
- XPES significantly increased microcracks in the middle third, while PTU did not.
- XPES induced more incomplete microcracks in the middle third compared to PTU.
Conclusions:
- No significant difference in microcrack formation between XPES and PTU in apical and cervical thirds.
- XPES instrumentation resulted in a higher incidence of incomplete microcracks in the middle root third.
- Careful consideration of instrument choice may be necessary to minimize dentinal damage.