Qualitative and quantitative image analysis of 16 cm wide-coverage computed tomography compared to new-generation
Liang Jin1, Yiyi Gao1, Yuqing Shan2
1Department of Radiology, Huadong Hospital (Affiliated to Fudan University), Shanghai, China.
Background:
Diagnostic quality of computed tomography (CT) images depends on numerous factors. Recently, two different modalities were introduced for coronary CT angiography (CCTA).
Objective:
This study aims to compare the performance of 16 cm wide-coverage detector CT (WDCT) using the snapshot freeze technique with a new-generation dual-source CT (DSCT) with 66 ms temporal resolution for CCTA.
Methods:
Total 101 patients with suspected coronary heart disease were enrolled. Of these, 50 and 51 patients were examined on WDCT and DSCT, respectively. CT values, image noise, signal-to-noise ratio, and contrast-to-noise ratio were measured. The image processing efficiency was recorded, followed by statistical comparison of diagnostic accuracy and radiation dose.
Results:
Ninety-nine patients (98.02%) had satisfactory diagnostic image quality. DSCT was significantly better than WDCT in terms of quantitative image quality, image processing efficiency, and qualitative analysis (P < 0.05). However, radiation dose was significantly lower on WDCT (P < 0.05) as compared to DSCT.
Conclusions:
Image processing efficiency and image quality of CCTA was higher on DSCT compared to WDCT due to the limitation of maximal tube current of WDCT.
More Related Videos
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT