Related Experiment Video
Updated: Dec 23, 2025

In Vivo Quantification of Hip Arthrokinematics during Dynamic Weight-bearing Activities using Dual Fluoroscopy
Published on: July 2, 2021
Biomechanical Evaluation of the Accuracy in Radiographic Assessment of Femoral Component Migration Measurement after
Mikayla Forness1,2, Zachary J Podoll1,2, Benjamin C Noonan2,3,4
1Sanford Sports Science Institute, Sanford Health, Fargo, ND.
Introduction:
Implant subsidence is one criteria utilized to monitor for prosthesis loosening after total hip arthroplasty (THA) with initial implant subsidence assessment often done utilizing plain radiographs. The specific aim of this study was to identify the most reliable references when using plain radiographs to establish an image magnification with the goals being easy to use, inexpensive, reliable, and accurate.
Methods:
Two femoral stem implants (stem lengths: 127 mm, 207 mm) were utilized to simulate hemiarthroplasty of the hip with composite femurs. Different combinations of femoral stem distances from the radiographic film (ODD), source-detector differences (SDD), hip rotation, and hip flexion were elected. Standardized anterior-posterior pelvis for each parameter combination setup were taken. Radiographic measurements (head diameter, stem length, stem seating length) were undertaken five times by three examiners. Radiographic image magnification factors were generated from two references (head diameter and stem length). Radiograph measurement reproducibility and stem seating length errors using these magnification factors were evaluated.
Results:
High level of repeated measurements reliability was found for head diameter (99 ± 0%) and stem length (90 ± 7%) measurements, whereas seating length measurements were less reliable (76 ± 6%). Stem length error using the femoral head magnification factor yielded 11% accuracy. Stem seating length error using both magnification factors were not reliable (< 7% accuracy). All parameters, except SDD, showed significant effect on calibrated measurement error.
Conclusion:
Current methods of assessing the implant subsidence after THA are inaccurate and unreliable. Clinicians should recognize these limitations and be cautious when diagnosing implant stability using plain radiographs alone.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
09:31Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018