RADIATION DOSE DURING PELVIC RADIOGRAPHY IN RELATION TO BODY MASS INDEX
Nika Zalokar1, Anja Resnik1, Nejc Mekiš1
1University of Ljubljana, Faculty of Health Sciences, Department of Medical Imaging and Radiotherapy, Zdravstvena pot 5, 1000 Ljubljana, Slovenia.
Abstract:
The purpose of this research was to investigate the impact of body mass index (BMI) on dose area product (DAP), effective dose (E), dose to the organs and image quality (IQ) on 200 patients referred to pelvic radiography. Patients were classified into three groups according to BMI: normal (<24.99), overweight (25.0-29.99) and obese (>30). The results showed 52% and 135% higher DAP for overweight and obese patients compared to normal-weight patients (p < 0.001). A 46 and 123% rise of E for overweight and obese patients compared to normal-weight patients (p < 0.001) was discovered. Overweight patients received 37% higher dose and obese patients 107% higher dose to the organs compared to normal-weight patients. There were no statistically significant differences between IQ, except between normal weight and overweight patients. A strong correlation (r = 0.733) was found between BMI and DAP and between BMI and E (r = 0.776).
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Drug Dosing: Obese Patients
Imaging Studies IV: Magnetic Resonance Imaging