Related Experiment Video
Updated: Dec 22, 2025

Selected Reaction Monitoring Mass Spectrometry for Absolute Protein Quantification
Published on: August 17, 2015
Evaluation, identification and impact assessment of abnormal internal standard response variability in regulated
Yunlin Fu1, Deborah Barkley1, Wenkui Li1
1Pharmacokinetic Sciences, Novartis Institutes for BioMedical Research, East Hanover, NJ 07936, USA.
Abstract:
Internal standard (IS) plays an important role in LC-MS bioanalysis by compensating for the variability of the analyte of interest in bioanalytical workflow. Due to the complexity of biological sample compositions and bioanalytical processes, a certain level of IS response variability across a run or a study is anticipated. However, an extensive variability may raise doubts to the accuracy of the measured results and also suggest nonoptimal analytical method. In this current paper, recent publications and guidelines regarding IS response in LC-MS bioanalysis were thoroughly reviewed with focus on the evaluation, identification and impact assessment of 'abnormal' IS response variability. A systematic decision tree was proposed to facilitate investigation into abnormal IS response variability after each run.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Data Validation
Key parameters for method validation include: