Related Experiment Video
Updated: Dec 22, 2025

Cooling an Optically Trapped Ultracold Fermi Gas by Periodical Driving
Published on: March 30, 2017
A simplified cryogenic optical resonator apparatus providing ultra-low frequency drift
Eugen Wiens1, Chang Jian Kwong1, Timo Müller2
1Institut für Experimentalphysik, Heinrich-Heine-Universität Düsseldorf, 40225 Düsseldorf, Germany.
Abstract:
A system providing an optical frequency with instability comparable to that of a hydrogen maser is presented. It consists of a 5 cm long, vertically oriented silicon optical resonator operated at temperatures between 1.5 K and 3.6 K in a closed-cycle cryostat with a low-temperature Joule-Thomson stage. We show that with a standard cryostat, a simple cryogenic optomechanical setup, and no active or passive vibration isolation, a minimum frequency instability of 2.5 × 10-15 at τ = 1500 s integration time can be reached. The influence of pulse-tube vibrations was minimized by using a resonator designed for low acceleration sensitivity. With reduced optical laser power and interrogation duty cycle, an ultra-low fractional frequency drift of -2.6 × 10-19/s is reached. At 3.5 K, the resonator frequency exhibits a vanishing thermal sensitivity and an ultra-small temperature derivative 8.5 × 10-12/K2. These are favorable properties that should lead to high performance also in simpler cryostats not equipped with a Joule-Thomson stage.