Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Influence of angle κ on visual and refractive outcomes after implantation of a diffractive trifocal intraocular lens
Nuria Garzón1, María García-Montero, Esther López-Artero
1From the IOA Madrid Innova Ocular (Garzón, López-Artero, Poyales); Department of Optica II, Complutense University, Madrid (Garzón, García-Montero); Clínica Oftalmológica Dr Gonzalo Muñoz, Valencia (Albarrán-Diego); Department of Ophthalmology (Oftalmar), Vithas Medimar International Hospital, Alicante (Pérez-Cambrodí); and Begitek, Innova Ocular (Illarramendi), Spain.
Purpose:
To evaluate changes in angle κ after the implantation of a trifocal intraocular lens (IOL) and to assess the postoperative outcomes of patients with different angle κ values.
Setting:
IOA Madrid Innova Ocular, Madrid, Spain.
Design:
Prospective trial.
Methods:
Sixty-three eyes from 63 patients who had bilateral implantation of a diffractive trifocal IOL (POD F, PhysIOL) were included. Pupil offset was used as the best estimate of angle κ and was measured using Pentacam (Oculus) preoperatively and at 3 months postoperatively. Postoperative refractive outcomes (sphere, cylinder, and manifest refraction spherical equivalent) and visual outcomes at far, intermediate, and near distance were assessed and compared between eyes with small pupil offset and eyes with large pupil offset. Quality of vision was assessed using a subjective questionnaire.
Results:
There was a significant decrease in pupil offset values postoperatively (mean: 0.197 ± 0.12 mm) compared with those preoperatively (mean: 0.239 ± 0.12 mm), with a mean decrease of -0.042 mm (P = .0002). The same significant decrease was found for both right and left eyes, when analyzed separately. No statistically significant difference was found in any of the refractive and visual acuity outcomes between eyes with small pupil offset and eyes with large pupil offset. The majority of patients (14 of 16) complaining of significant halos had eyes with small pupil offset.
Conclusions:
Large pupil offset did not negatively affect visual and refractive outcomes. The tolerance to larger pupil offset might be due to the IOL optical design, with the first diffractive ring being larger than other commonly used multifocal IOLs. More studies comparing various diffractive IOL models would be useful to confirm such hypothesis.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview