Traceable point-like 68Ge/68Ga source with a spherically symmetric positron absorber for PET scanners
Tomoyuki Hasegawa1,2, Mio Okamoto3, Takahiro Yamada4
1School of Allied Health Sciences, Kitasato University, 1-15-1, Kitasato, Minamiku, Sagamihara, Kanagawa, 252-0373, Japan. hasegawa@kitasato-u.ac.jp.
Abstract:
A novel traceable point-like 68Ge/68Ga source was developed for calibration of positron emission tomography (PET) scanners. The source was equipped with a spherically symmetric acrylic positron absorber. The physical characteristics of photons emitted from the point-like source were evaluated by Monte Carlo simulation, considering possible geometrical uncertainties. The calibration factor of a clinical PET/CT scanner was evaluated using four manufactured point-like sources as a practical application of the point-like source. The results were consistent with that determined by the conventional cross-calibration method. The traceable point-like 68Ge/68Ga source is expected to be a simple and practical tool for determining the calibration factor and evaluating the physical characteristics of PET scanners.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET