Optimization and validation of patient-based real-time quality control procedure using moving average and average of
Chao Song1, Jun Zhou2, Jun Xia3
1Zhejiang Center for Clinical Laboratories, Zhejiang Provincial People's Hospital, Hangzhou, China.
Background:
We have designed a patient-based real-time quality control (PBRTQC) procedure to detect analytical shifts and review analytical trends of measurement procedures.
Methods:
All the nine months' patient results of total thyroxine (TT4), total triiodothyronine (TT3), free thyroxine (FT4), free triiodothyronine (FT3), and thyrotropin (TSH) measured by three identical analyzers were divided into three groups according to the source of inpatient patients, outpatient patients, and healthy people. The data in each group were truncated by optimized Box-Plot method and normalized by Box-Cox method if necessary. The z-score charts of internal quality control (IQC) samples' results and PBRTQC data were drawn by IQC levels and groups, respectively. The analytical shifts and analytical trends were detected by multi-rules of 2-2S rules and moving average rules. The performances of PBRTQC were compared with the BIQC in which IQC samples were measurand only once per day at the beginning of the analytical batch. Twelve quality control cases were listed to validate the performances.
Results:
All the five analytes presented normal distributions when the parameter n of Box-Plot method was 1.2. The percentages of excluded data ranged from 2.9% to 11.6%. 31 and 14 rejections triggered in PBRTQC and BIQC, respectively. 96.8% of the shift rejections in PBRTQC were trend-related shifts and calibration-related shifts, while the proportion was 85.7% in BIQC but 78.6% of the shift rejections in TSH. 25.7% and 8.6% of 105 calibration events which caused analytical shifts were detected by PBRTQC and BIQC, respectively. However, the performance of PBRTQC was not well in TSH because of its large coefficient of variation.
Conclusions:
The optimized PBRTQC is high efficiency than BIQC in detecting analytical shifts, trends, and calibration events. The PBRTQC can be used as a low-cost supplementary procedure to IQC every day, especially at the end of the analytical batch on that day when the within-individual biological variation of analyte is not larger than its coefficient of variation in IQC. Further optimization and validation of PBRTQC are still needed.
Related Concept Videos
Therapeutic Drug Monitoring: Drug Analysis Methods
Therapeutic Drug Monitoring: Overview and Classification
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include:
Therapeutic Drug Monitoring: Affecting Factors