Validation and applicability of an alternative method for dialysis water and dialysate quality analysis
Gabriela Corrêa Carvalho1, Adriana Bugno2, Adriana Aparecida Buzzo Almodovar2
1Universidade de São Paulo, Faculdade de Ciências Farmacêuticas, Departamento de Farmácia, São Paulo, SP, Brasil.
Introduction:
In hemodialysis, patients are exposed to a large volume of water, which may lead to fatal risks if not meeting quality standards. This study aimed to validate an alternative method for monitoring microbiological quality of treated water and assess its applicability in dialysis and dialysate analysis, to allow corrective actions in real-time.
Methods:
Validation and applicability were analyzed by conventional and alternative methods. For validation, E. coli standard endotoxin was diluted with apyrogenic water in five concentrations. For the applicability analysis, treated water for dialysis was collected from different points in the treatment system (reverse osmosis, drainage canalization at the storage tank bottom, reuse, and loop), and dialysate was collected from four machines located in different rooms in the hemodialysis sector.
Results:
The validation results were in accordance with the Brazilian Pharmacopoeia acceptance criteria, except for the last two concentrations analyzed. In addition, the ruggedness criterion performed under the US Pharmacopoeia was in agreement with the results.
Discussion:
A limiting factor in the applicability analysis was the absence of the endotoxin maximum permitted level in dialysate by the Brazilian legislation. When comparing the analysis time, the alternative method was more time-consuming than the conventional one. This suggests that the alternative method is effective in the case of few analyses, that is, real-time analyses, favoring corrective actions promptly. On the other hand, it does not support the implementation of the alternative method in a laboratory routine due to the high demand for analyses.
More Related Videos
11:43Dissolved Solute Sampling Across an Oxic-Anoxic Soil-Water Interface Using Microdialysis Profilers
Published on: March 24, 2023
12:11A Modified EPA Method 1623 that Uses Tangential Flow Hollow-fiber Ultrafiltration and Heat Dissociation Steps to Detect Waterborne Cryptosporidium and Giardia spp.
Published on: July 9, 2012
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Data Validation
Key parameters for method validation include:
Testing Water Quality
Quality of Water
Hemodialysis II: Procedure and Complications