Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors
Jesús Pérez-Gracia1, Alejandra Varea1, Jorge Ares1
1Departamento de Física Aplicada, Universidad de Zaragoza, Zaragoza, Spain.
Purpose:
To evaluate and compare the effect of misalignment and tilt on the optical performance of different aspheric intraocular lens (IOL) designs.
Methods:
Three aspheric IOLs with a different quantity of spherical aberration (SA) have been designed and the effect of IOL misalignment and tilt on the imaging quality of an eye model has been numerically assessed using a commercial optical design software. The prototypes have been manufactured by lathe turning and tested in vitro using the same optical bench (PMTF, Lambda-X) that complies with International Organization for Standardization standard 11979-2 requirements. Image quality was evaluated from the modulation transfer functions (MTFs), through-focus modulation transfer functions (TF-MTFs), root mean square (RMS) values of defocus, astigmatism and coma, and images of the United States Air Force (USAF) target were taken. A comparison with the optical performance of spherical IOLs has also been performed.
Results:
Intraocular lens misalignment and tilt increased wavefront aberrations; the effect of misalignment on root mean square (RMS) astigmatism and coma was positively correlated with the spherical aberration of the IOL. Aberration-free IOLs showed the highest MTF for all misalignment values and for IOLs with negative SA correction the MTF decays below 0.43 when they are decentered 0.50 mm.
Conclusions:
Aspherical IOLs are more sensitive than spherical IOLs to misalignment or tilt, depending on their SA correction. The optical degradation caused by IOL misalignment had a greater effect on IOL designs with a higher amount of negative spherical aberration. In contrast, the effect of tilt on the optical performance was less sensitive to the IOL design.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Properties of Enantiomers and Optical Activity