A new method for measuring the pneumotachometer characteristics using a syringe
Related Experiment Video
Updated: Oct 16, 2026

A Reversible, Non-invasive Method for Airway Resistance Measurements and Bronchoalveolar Lavage Fluid Sampling in Mice
Published on: April 13, 2010
1Department of Internal Medicine, Kanazawa Medical University, Ishikawa, Japan.
Abstract:
A new method for determining the nonlinear characteristics of pneumotachometers was developed. This method consists of a computer algorithm analyzing unsteady flows generated by a syringe. We removed the possible influence of the frequency characteristics of the measuring system by inserting an acoustic low pass filter in the path of gas flow. Two types of Fleisch pneumotachometers were tested. For a given volume of 3 l, the error of measured volume is 12.5% in uncorrected series and 0.2% in the corrected.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...