Objectives:
The implantation of flow diverters, or stents in general, necessitates the use of dual anti-platelet treatment with typical regimes including aspirin and a P2Y12 inhibitor. This carries an inherent risk of haemorrhage. We sought to compare the thrombogenicity of the anti-thrombogenic p48 hydrophilic polymer coating compared to the standard uncoated p48 flow diverter using an in vitro thrombogenicity assay.
Methods:
To evaluate the thrombin generation influenced by the different stent types the stents were placed in wells of a 24-well plate with the addition of plasma from healthy volunteers the thrombin calibrator respectively the PPP-reagent was added. Subsequently, the thrombin substrate was added and the thrombin generation was analysed every 60 s using a thrombinoscope. The assay is calibrated using samples containing a known amount of active thrombin in PPP. Thrombin activity is proportional to the change in fluorescence.
Results:
The p48 hydrophilic polymer coating shows a significantly lower peak thrombin concentration (1.13 ± 0.21 vs. 1.41 ± 0.22) and longer time to peak thrombin concentration (0.96 ± 0.04 vs. 0.74 ± 0.07) compared to the uncoated p48 device (p < 0.01). The responses of the p48 hydrophilic polymer coating were similar to that of the negative control.
Conclusion:
The hydrophilic polymer coating surface modification significantly reduces the thrombogenicity of the p48 flow diverter. These results corroborate the findings from previous in vitro studies.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Antiplatelet Drugs: Prostaglandin Synthesis, P2Y12 and Glycoprotein IIb/IIIa Inhibitors
Prostaglandin synthesis inhibitors, exemplified by the widely known aspirin, wield their power by irreversibly acetylating...