Clinical Performance of Flash Glucose Monitoring System in Patients with Liver Cirrhosis and Diabetes Mellitus
Dalila Costa1,2, Joana Lourenço3, Ana Margarida Monteiro4
1Life and Health Sciences Research Institute (ICVS), School of Medicine, University of Minho, Braga, Portugal. dc.dalilacosta@gmail.com.
Abstract:
Flash glucose monitoring system (FGMS) is an improved subset of continuous glucose monitoring with a recognized effectiveness on glycemic control, though validation in patients with Liver Cirrhosis (LC) is lacking. To evaluate the accuracy of FGMS in patients with Type 2 Diabetes Mellitus (DM) and LC, a prospective, case-control study was performed in 61 ambulatory patients with LC and DM (LC group, n = 31) or DM (Control group, n = 30). During 14 days, patients performed 4 assessments per day of self-monitoring of blood glucose (SMBG, reference value) followed by FGMS scanning. There were 2567 paired SMBG and FGMS values used in the accuracy analysis, with an overall mean absolute relative difference (MARD) of 12.68% in the LC group and 10.55% in the control group (p < 0,001). In patients with LC, the percentage of readings within Consensus Consensus Error Grid analysis Zone A and A + B were 80.36% and 99,26%, respectively. Sensor clinical accuracy was not affected by factors such as body mass index, age, gender, Child-Pugh score or edematoascitic decompensation. This is the first study to approach FGMS clinical accuracy in LC, revealing a potential usability of this system to monitor glycemic control in this population.
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Diabetes Mellitus: Type 2 and Gestational