The influence of SBP amplification on the accuracy of form-factor-derived mean arterial pressure
Related Experiment Video
Updated: Dec 22, 2025

Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
Published on: December 2, 2014
Martin G Schultz1, Dean S Picone1, Mathew K Armstrong1
1Menzies Institute for Medical Research, College of Health and Medicine, University of Tasmania.
Objectives:
Accurate assessment of mean arterial pressure (MAP) is crucial in research and clinical settings. Measurement of MAP requires not only pressure waveform integration but can also be estimated via form-factor equations incorporating peripheral SBP. SBP may increase variably from central-to-peripheral arteries (SBP amplification), and could influence accuracy of form-factor-derived MAP, which we aimed to determine.
Methods:
One hundred and eighty-eight patients (69% men, age 60 ± 10 years) undergoing coronary angiography had intra-arterial pressure measured in the ascending aorta, brachial and radial arteries. Reference MAP was measured by waveform integration, and form-factor-derived MAP using 33 and 40% form-factors.
Results:
Reference MAP decreased from the aorta to the brachial (-0.7 ± 4.2 mmHg) and radial artery (-1.7 ± 4.8 mmHg), whereas form-factor-derived MAP increased (33% form-factor 1.1 ± 4.2 and 1.7 ± 4.7 mmHg; 40% form-factor 0.9 ± 4.8 and 1.4 ± 5.4 mmHg, respectively). Form-factor-derived MAP was significantly different to reference aortic MAP (33% form-factor -2.5 ± 4.6 and -1.6 ± 5.8, P < 0.001; 40% form-factor 2.5 ± 5.0 and 3.9 ± 6.4 mmHg, P < 0.001, brachial and radial arteries, respectively), with significant variation in the brachial form-factor required (FFreq) to generate MAP equivalent to reference aortic MAP (FFreq range 20-57% brachial; 17-74% radial). Aortic-to-brachial SBP amplification was strongly related to brachial FFreq (r = -0.695, P < 0.001). The 33% form-factor was most accurate with high aortic-to-brachial SBP amplification (33% form-factor MAP vs. reference aortic MAP difference 0.06 ± 3.93 mmHg, P = 0.89) but overestimated reference aortic MAP with low aortic-to-brachial SBP amplification (+5.8 ± 4.6 mmHg, P < 0.001). The opposite was observed for the 40% form-factor.
Conclusion:
Due to variable SBP amplification, estimating MAP via form-factors produces nonphysiological inaccurate values. These findings have important implications for accurate assessment of MAP in research and clinical settings.
More Related Videos
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure