Related Experiment Video
Updated: Dec 22, 2025

Treatment of Platelet Products with Riboflavin and UV Light: Effectiveness Against High Titer Bacterial Contamination
Published on: August 24, 2015
Hepatitis E virus is effectively inactivated in platelet concentrates by ultraviolet C light
Dimas Praditya1,2, Martina Friesland3, Ute Gravemann4
1Department of Molecular and Medical Virology, Faculty of Medicine, Ruhr University Bochum, Bochum, Germany.
Background And Objectives:
As previous investigations have shown, THERAFLEX UV-Platelets, a UVC-based pathogen inactivation (PI) system, is effective against non-enveloped transfusion-relevant viruses such as hepatitis A virus (HAV), which are insensitive to most PI treatments for blood products. This study investigated the PI efficacy of THERAFLEX UV-Platelets against HEV in platelet concentrates (PCs).
Materials And Methods:
Buffy coat-derived PCs in additive solution were spiked with cell culture-derived HEV and treated with the THERAFLEX UV-Platelets system using various doses of UVC (0·05, 0·10, 0·15 and 0·20 (standard) J/cm2 ). Titres of infectious virus in pre- and post-treatment samples were determined using a large-volume plating assay to improve the detection limit of the virus assay.
Results:
THERAFLEX UV-Platelets dose-dependently inactivated HEV in PCs. The standard UVC dose inactivated the virus to below the limit of detection, corresponding to a mean log reduction of greater than 3·5.
Conclusion:
Our study demonstrates that the THERAFLEX UV-Platelets system effectively inactivates HEV in PCs.
More Related Videos
12:40Preparation and Pathogen Inactivation of Double Dose Buffy Coat Platelet Products using the INTERCEPT Blood System
Published on: December 7, 2012
05:49Procoagulant Platelet Characterization by Measuring Phosphatidylserine Exposure and Microvesicle Release from Human Purified Platelets