Related Experiment Video
Updated: Dec 21, 2025

Recording Large-scale Neuronal Ensembles with Silicon Probes in the Anesthetized Rat
Published on: October 19, 2011
Recording local field potential and neuronal activity with tetrodes in epileptic patients
Elodie Despouy1, Jonathan Curot2, Leila Reddy3
1Centre de Recherche Cerveau et Cognition, Université de Toulouse, Université Paul Sabatier Toulouse, Toulouse F-31330, France; Centre National de la Recherche Scientifique CerCo, Toulouse F-31052, France; DIXI Medical, Chaudefontaine F-25640 France.
Background:
Recordings with tetrodes have proven to be more effective in isolating single neuron spiking activity than with single microwires. However, tetrodes have never been used in humans. We report on the characteristics, safety, compatibility with clinical intracranial recordings in epileptic patients, and performance, of a new type of hybrid electrode equipped with tetrodes.
New Method:
240 standard clinical macroelectrodes and 102 hybrid electrodes were implanted in 28 patients. Hybrids (diameter 800 microm) are made of 6 or 9 macro-contacts and 2 or 3 tetrodes (diameter 70-80 microm).
Results:
No clinical complication or adverse event was associated with the hybrids. Impedance and noise of recordings were stable over time. The design enabled multiscale spatial analyses that revealed physiopathological events which were sometimes specific to one tetrode, but could not be recorded on the macro-contacts. After spike sorting, the single-unit yield was similar to other hybrid electrodes and was sometimes as high as >10 neurons per tetrode.
Comparison With Existing Method(S):
This new hybrid electrode has a smaller diameter than other available hybrid electrodes. It provides novel spatial information due to the configuration of the tetrodes. The single-unit yield appears promising.
Conclusions:
This new hybrid electrode is safe, easy to use, and works satisfactorily for conducting multi-scale seizure and physiological analyses.

