Two-dimensional UTE overview imaging for dental application
Kilian Stumpf1, Elena Kaye2, Jan Paul1,3
1Department of Internal Medicine II, Ulm University Medical Center, Ulm, Germany.
Purpose:
To investigate the applicability of a 2D-UTE half-pulse sequence for dental overview imaging and the detection of signal from mineralized dental tissue and caries lesions with ultra-short as an efficient alternative to 3D sequences.
Methods:
A modified 2D-UTE sequence using 240-µs half-pulses for excitation and a reduction of the coil tune delay from the manufacturer preset value allowed for the acquisition of in vivo dental images with a TE of 35 µs at 1.5T. The common occurrence of out-of-slice signal for half-pulse sequences was avoided by applying a quadratic-phase saturation pulse before each half-RF excitation. A conventional 2D-UTE sequence with a TE of 750 µs, using slice selection rephasing, was used for comparison.
Results:
Quadratic phase saturation pulses adequately improve the slice profile of half-pulse excitations for dental imaging with a surface coil. In vivo images and SNR measurements show a distinct increase in signal in ultrashort tissues for the proposed 2D-UTE half-pulse sequence compared with a 2D-UTE sequence using conventional slice selection, leading to an improved detection of caries lesions.
Conclusion:
The proposed pulse sequence enables the acquisition of in vivo images of a comprehensive overview of bone structures and teeth of a single side of the upper and lower jaw and signal detection from mineralized dental tissues in clinically acceptable scan times.
More Related Videos
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):