Dosimetric evaluation of PASSAG-U polymer gel dosimeter: Dependence of dose rate and photon energy
Bagher Farhood1, Kamal Mohammadi Asl2, Mostafa Sarvizadeh3
1Department of Medical Physics and Radiology, Faculty of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran.
Objective:
Several physical factors such as dose rate and photon energy may change response and sensitivity of polymer gel dosimeters. This study aims to evaluate the R2-dose response and sensitivity dependence of PASSAG-U gel dosimeters with 3% and 5% urea on dose rate and photon energy.
Materials And Methods:
The PASSAG-U gel dosimeters were prepared under normal atmospheric conditions. The obtained gel dosimeters were irradiated to different dose rates (100, 200, and 300 cGy/min) and photon energies (6 and 15 MV). Finally, responses (R2) of the PASSAG-U gel dosimeters with 3% and 5% urea were analyzed by MRI technique at 1, 10, 14 days after the irradiation process.
Results:
The findings showed that the R2-dose responses of PASSAG-U gel dosimeters with 3% and 5% urea do not vary under the differently evaluated dose rates and photon energies. The R2-dose sensitivity of PASSAG-U polymer gel dosimeter with 3% urea does not change under the differently evaluated dose rates and photon energies, but it changes for PASSAG-U polymer gel dosimeter with 5% urea. The dose resolution values ranged from 0.20 to 0.86 Gy and from 0.27 to 2.20 Gy for the PASSAG-U gel dosimeter with 3% and 5% urea for the different dose rates and photon energies, respectively. Furthermore, it was revealed that the R2-dose response and sensitivity dependence of PASSAG-U gel dosimeters with 3% and 5% urea on dose rate and photon energy can vary over post irradiation time.
Conclusions:
The study results demonstrated that dosimetric characteristics (dependence of dose rate and photon energy, and dose resolution) of PASSAG-U gel dosimeter with 3% were better than those of PASSAG-U gel dosimeter with 5% urea.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Biological Effects of Radiation
Dose Size and Dosing Frequency: Determination Methods
Dose-Response Relationship: Overview
Dose-Response Relationship: Potency and Efficacy
Dosage Regimens: Partial Pharmacokinetic Parameters