Improved image quality of temporal bone CT with an ultrahigh-resolution CT scanner: clinical pilot studies
Arisa Ohara1, Haruhiko Machida2, Hisae Shiga1
1Department of Radiology, Kyorin University Faculty of Medicine, 6-20-2, Shinkawa, Mitaka-shi, Tokyo, 181-8611, Japan.
Purpose:
Ultrahigh-resolution CT (UHRCT) with slice collimation of 0.25 mm × 160 and matrix size of 1024 × 1024 has become clinically available. We compared the image quality of temporal bone CT (TBCT) between UHRCT and conventional multidetector CT (MDCT).
Materials And Methods:
We retrospectively enrolled 20 patients who underwent TBCT by MDCT (matrix size, 512 × 512) and subsequently by UHRCT (matrix size, 1024 × 1024). Two independent reviewers subjectively graded delineation of normal stapes, oval window, facial nerve canal, incudostapedial joint, and tympanic tegmen. We also quantified image noise in the cerebellar hemisphere. Between MDCT and UHRCT, we compared mean subjective grades using the Wilcoxon signed-rank test and the image noise using paired t test.
Results:
Grades were significantly higher with UHRCT than with MDCT for all the anatomies (P < 0.001), whereas noise was significantly higher with UHRCT than with MDCT (P = 0.002).
Conclusion:
For TBCT, UHRCT shows better delineation of the fine anatomical structures compared with MDCT.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography