Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Analytical performance evaluation of Sysmex CS-2500 and Stago STA Compact
Ozgur M Yis1, Guler Bugdayci1, Mine B Pehlivan1
1Department of Medical Biochemistry, Faculty of Medicine, Bolu Abant Izzet Baysal University.
The Sysmex Coagulation System-2500 (CS-2500) demonstrates acceptable performance for prothrombin time (PT) and activated partial thromboplastin time (aPTT) tests. Minor differences were observed compared to another analyzer, suggesting potential for future standardization.
Area of Science:
- Clinical diagnostics
- Biomedical engineering
- Laboratory medicine
Background:
- Coagulation testing is crucial for diagnosing and managing bleeding disorders.
- Automated analyzers improve efficiency and accuracy in clinical laboratories.
- The Sysmex Coagulation System-2500 (CS-2500) is a new fully automated analyzer utilizing optical detection.
Purpose of the Study:
- To evaluate the performance characteristics of the CS-2500 analyzer.
- To assess the accuracy and precision of PT and aPTT measurements.
- To compare the CS-2500 with an existing coagulometer (STA-Compact).
Main Methods:
- Performance evaluation of the CS-2500 for prothrombin time (PT) and activated partial thromboplastin time (aPTT).
- Precision, accuracy, and reference range verification studies were conducted.
- Method comparison using Passing-Bablok regression analysis against the STA-Compact analyzer.
Main Results:
- Excellent precision was observed, with coefficient of variation < 3.5% for PT-INR and aPTT.
- Good agreement between CS-2500 and STA-Compact for PT-INR (r=0.962) and aPTT (r=0.944).
- Bias was < 9.9% in accuracy studies, and manufacturer reference ranges were acceptable.
Conclusions:
- The CS-2500 demonstrates acceptable performance for PT and aPTT testing.
- Minor discrepancies between analyzers were noted, highlighting the need for standardization.
- Further studies are recommended to harmonize results across different coagulation systems.