Related Experiment Videos
Pressure resistance of breast implants as a function of implantation time
J H van Rappard1, G J Sonneveld, R van Twisk
1Department of Plastic and Reconstructive Surgery, University of Nijmegen, the Netherlands.
Annals of Plastic Surgery
|December 1, 1988
Abstract:
Using a relatively simple test, breaking pressures of 50 gel-filled breast implants that had to be removed have been evaluated. These pressures values were compared with forces generated in executing closed capsulotomy (nutcracker method). It is shown that there is a negative correlation between pressure resistance of breast implants and duration of implantation. After an implantation time of more than five years, our advice is to perform closed capsulotomy if done in a very gentle way, and implant rupture then certainly has to be taken into account.