Accuracy and Trending Ability of Blood Pressure and Cardiac Index Measured by ClearSight System in Patients With
Hiroko Kanazawa1, Takuma Maeda2, Erika Miyazaki1
1Department of Anesthesiology, National Cerebral and Cardiovascular Center, Osaka, Japan.
Objectives:
To investigate the accuracy and trending ability of ClearSight (Edwards Lifesciences, Irvine, CA) in patients with reduced ejection fraction (<55%) undergoing off-pump coronary artery bypass graft (CABG) surgery by comparing the ClearSight-derived cardiac index (CICS) with the cardiac index measured with thermodilution using a pulmonary artery catheter. In addition, the accuracy and trending ability of ClearSight for blood pressure measurement was investigated by comparing the mean arterial pressure (MAP) derived by ClearSight (MAPcs) with invasive intra-arterial pressure.
Design:
Prospective clinical study.
Design:
Cardiac surgery operating room in a single cardiovascular center.
Participants:
The study comprised 20 patients who underwent elective CABG surgery.
Interventions:
MAP and cardiac index were measured simultaneously at 6 time points intraoperatively. Trending ability was investigated at the following 2 points: (1) before and after placing the patient in the Trendelenburg position and (2) before and after atrial pacing with a targeted heart rate increase of 20%.
Measurements And Main Results:
Bland-Altman analysis showed that the percentage error between CICS and the cardiac index measured with thermodilution was 40.2% and the percentage error between MAPcs and MAP was 24.6%. Four-quadrant plot analysis showed that the tracking ability of CICS with the Trendelenburg position and atrial pacing was below the good trending ability cutoff (92%). However, the concordance rate of the 4-quadrant plot analysis showed a good trending ability for MAPcs. The polar plot analysis showed the same trend.
Conclusions:
CICS was not sufficiently accurate in patients with reduced ejection fraction undergoing off-pump CABG surgery. However, ClearSight was clinically acceptable for MAP regarding its accuracy and trending ability in patients with reduced ejection fraction.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Assessing Blood Pressure