Related Experiment Video
Updated: Dec 21, 2025

Author Spotlight: Improved Imaging for Neovascular Development in Congenital Heart Disease Research
Published on: April 26, 2024
Long-Term Venovenous Connection for Extracorporeal Carbon Dioxide Removal (ECCO2R)-Numerical Investigation of the
N B Steuer1, K Hugenroth2, T Beck2
1Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany. steuer@ame.rwth-aachen.de.
Purpose:
Currently used cannulae for extracorporeal carbon dioxide removal (ECCO2R) are associated with complications such as thrombosis and distal limb ischemia, especially for long-term use. We hypothesize that the risk of these complications is reducible by attaching hemodynamically optimized grafts to the patient's vessels. In this study, as a first step towards a long-term stable ECCO2R connection, we investigated the feasibility of a venovenous connection to the common iliac veins. To ensure its applicability, the drainage of reinfused blood (recirculation) and high wall shear stress (WSS) must be avoided.
Methods:
A reference model was selected for computational fluid dynamics, on the basis of the analysis of imaging data. Initially, a sensitivity analysis regarding recirculation was conducted using as variables: blood flow, the distance of drainage and return to the iliocaval junction, as well as the diameter and position of the grafts. Subsequently, the connection was optimized regarding recirculation and the WSS was evaluated. We validated the simulations in a silicone model traversed by dyed fluid.
Results:
The simulations were in good agreement with the validation measurements (mean deviation 1.64%). The recirculation ranged from 32.1 to 0%. The maximum WSS did not exceed 5.57 Pa. The position and diameter of the return graft show the highest influence on recirculation. A correlation was ascertained between recirculation and WSS. Overall, an inflow jet directed at a vessel wall entails not only high WSS, but also a flow separation and thereby an increased recirculation. Therefore, return grafts aligned to the vena cava are crucial.
Conclusion:
In conclusion, a connection without recirculation could be feasible and therefore provides a promising option for a long-term ECCO2R connection.
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Hemodialysis II: Procedure and Complications
Overview of Systemic Veins
The coronary sinus, the heart's principal vein, resides in the coronary sulcus on the heart's posterior aspect. This broad venous channel receives nearly all venous blood from the myocardium, the heart muscle. It is fed by three primary veins: the great cardiac vein, the...
Hemodialysis I: Introduction
Varicose Veins II: Diagnostic Studies and Interprofessional Care