New sensitive method for HEPES quantification in 68Ga-radiopharmaceuticals
I F Antunes1, G M Franssen2, R Zijlma3
1Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands. i.farinha.antunes@umcg.nl.
EJNMMI Radiopharmacy and Chemistry
|May 16, 2020
Summary
A new HPLC method accurately quantifies 2-[4-(2-hydroxyethyl)piperazin-1-yl]ethanesulfonic acid (HEPES) in 68Ga-radiopharmaceuticals, unlike less sensitive TLC methods. This ensures patient safety by preventing excessive HEPES doses.
Area of Science:
- Nuclear Medicine
- Radiopharmaceutical Chemistry
- Analytical Chemistry
Background:
- GMP-certified 68Ga generators have increased 68Ga-radiopharmaceutical use.
- 2-[4-(2-hydroxyethyl)piperazin-1-yl]ethanesulfonic acid (HEPES) buffer improves radiosynthesis but is an impurity.
- European Pharmacopoeia (Ph Eur) limits HEPES to 200 μg/VInjected for radiopharmaceuticals like edotreotide.
Purpose of the Study:
- To evaluate the sensitivity of the Ph Eur thin-layer chromatography (TLC) method for HEPES quantification.
- To develop and validate a new high-performance liquid chromatography (HPLC) method for HEPES quantification in 68Ga-radiopharmaceuticals.
- To compare the new HPLC method with the Ph Eur TLC method and assess HEPES levels in various 68Ga-radiopharmaceuticals from different institutes.
Main Methods:
- Tested the Ph Eur TLC method for HEPES detection in 68Ga-radiopharmaceuticals.
- Developed and validated a modified TLC method and a new HPLC method for HEPES quantification.
- Analyzed 68Ga-radiopharmaceutical samples from multiple institutions using the developed methods.
Main Results:
- The Ph Eur TLC method failed to detect HEPES within 4 minutes of iodine incubation.
- The modified TLC method required 2 hours for visible spots, only at concentrations >1 mg/mL.
- The HPLC method achieved a limit of quantification (LOQ) of 3 μg/mL and a limit of detection (LOD) of 1 μg/mL, detecting excessive HEPES in [68Ga]Ga-NODAGA-Exendin samples.
Conclusions:
- The Ph Eur and modified TLC methods are insufficiently sensitive for quality control (QC) of HEPES in 68Ga-radiopharmaceuticals.
- The new HPLC method is sensitive, quantitative, reproducible, and suitable for QC release.
- This HPLC method identified 68Ga-radiopharmaceutical batches exceeding the pharmacopoeial HEPES limit, enabling dose adjustment for patient safety.