Dual-Energy CT-Based Iodine Quantification in Liver Tumors - Impact of Scan-, Patient-, and Position-Related Factors
Cynthia Schmidt1, Bettina Baessler1, Dominik Nakhostin1
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Rämistrasse 100, Zurich 8091, Switzerland.
Rationale And Objectives:
To quantify the contribution of lesion location and patient positioning, dual-energy approach, patient size, and radiation dose to the error of dual-energy CT-based iodine quantification (DECT-IQ) in liver tumors.
Materials And Methods:
A phantom with four liver lesions (diameter 15 mm; iodine concentration 0-5 mgI/mL) and two sizes was used. One lesion emulated a subdiaphragmatic lesion. Both sizes were imaged in dual-energy mode on (1) a dual-source DECT (DS-DE) at 100/Sn150 kV and (2) a single-source split-filter DECT (SF-DE) at AuSn120 kV at two radiation doses (8 and 12 mGy). Scans were performed at seven different vertical table positions (from -6 to + 6 cm from the gantry isocenter). Iodine concentration was repeatedly measured and absolute errors (errorabs) were calculated. Errors were compared using robust repeated-measures ANOVAs with post-hoc comparisons. A linear mixed effect model was used to determine the factors influencing the error of DECT-IQ.
Results:
The linear mixed effect models showed that errors were significantly influenced by DECT approach, phantom size, and lesion location (all p < 0.001). The impact of lesion location on the error was stronger in SF-DE compared to DS-DE. Radiation dose did not significantly influence error (p = 0.22). When averaged across all setups, errorabs was significantly higher for SF-DE (2.08 ± 1.92 mgI/mL) compared to DS-DE (0.37 ± 0.29 mgI/mL) (all p < 0.001). Artefacts were found in the subdiaphragmatic lesion for SF-DE with significantly increased errorabs compared to DS-DE (p < 0.001). Errorabs was significantly higher in the large compared to the medium phantom for DS-DE (0.30 ± 0.23 mgI/mL vs. 0.43 ± 0.33 mgI/mL) and SF-DE (1.68 ± 1.99 vs. 2.36 ± 1.81 mgI/mL) (p < 0.001).
Conclusion:
The dual-energy approach, patient size, and lesion location modified by patient position significantly impacted DECT-IQ in simulated liver tumors.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT