Comparison of Cochlear Implant Magnets and Their MRI Artifact Size
I Todt1, R Guerkov1, H B Gehl2
1Department of Otolaryngology, Head and Neck Surgery, Bielefeld University, Campus Mitte, Bielefeld, Germany.
Introduction:
Recent developments regarding cochlear implant magnets (e.g., a bipolar diametral magnet) and refined surgical techniques (e.g., implant positioning) have had a significant impact on the relation between cochlear implants and MRIs, making the reproducible visibility of cochlea and IAC possible. MRI scanning has changed from a contraindication to a diagnostic tool. Magnet artifact size plays a central role in the visual assessment of the cochlea and IAC.
Objective:
The aim of this study is to compare the CI magnet-related maximum artifact sizes of various cochlea implant systems.
Materials And Methods:
We performed an in vivo measurement of MRI artifacts at 1.5 and 3 Tesla with three cochlear implant magnet systems (AB 3D, Medel Synchrony, and Oticon ZTI). The implant, including the magnet, was positioned with a head bandage 7.0 cm and 120° from the nasion, external auditory canal. We used a TSE T2w MRI sequence on the axial and coronal plains and compared the artifacts in two volunteers for each tesla strength.
Results:
Intraindividual artifact size differences between the three magnets are smaller than interindividual maximum artifact size differences. 3 T MRI scans, in comparison to 1.5 T MRI scans, show a difference between soft artifact areas.
Conclusion:
We observed no major difference between maximum implant magnet artifact sizes of the three implant magnet types.
More Related Videos
08:57Measuring the Influence of Magnetic Vestibular Stimulation on Nystagmus, Self-Motion Perception, and Cognitive Performance in a 7T MRT
Published on: March 3, 2023
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI