TO USE OR NOT USE PATIENT SHIELDING ON PREGNANT WOMEN UNDERGOING CT PULMONARY ANGIOGRAPHY: A PHANTOM STUDY
Dino Begano1, Marcus Söderberg2,3, Anetta Bolejko1,4
1Department of Medical Imaging and Physiology, Skåne University Hospital, Malmö SE-205 02, Sweden.
Pregnancy increases the risk of pulmonary embolism. Computed tomography pulmonary angiography (CTPA) is used for diagnosis. CT generates ionising radiation, and thus, abdominal shielding may be used. This phantom study investigated the effects of patient shielding and scan length reduction on the fetal and maternal ionising radiation dose from CTPA. The absorbed dose to the fetus was measured using thermoluminescent dosemeters. Estimated effective doses to the pregnant patient were based on the dose-length products. Shielding increased both the effective dose to the patient by 47% and the mean absorbed dose to the fetus (0.10 vs. 0.12 mGy; p < 0.001) compared with unshielded standard CTPA, as it affected the automatic exposure control. Shielded short CTPA marginally lowered only the mean fetal absorbed dose (0.03 vs. 0.02 mGy; p = 0.018). Shortening the scan reduced the fetal absorbed dose most effectively by 70% (0.10 vs. 0.03 mGy; p = 0.006), compared with the standard unshielded scan. Shielding modestly reduces fetal radiation dose but may compromise automatic exposure control, possibly increasing the maternal and fetal radiation dose. Shortening the scan is beneficial, assuming anatomical coverage is secured.
Pregnancy increases the risk of pulmonary embolism. Computed tomography pulmonary angiography (CTPA) is used for diagnosis. CT generates ionising radiation, and thus, abdominal shielding may be used. This phantom study investigated the effects of patient shielding and scan length reduction on the fetal and maternal ionising radiation dose from CTPA. The absorbed dose to the fetus was measured using thermoluminescent dosemeters. Estimated effective doses to the pregnant patient were based on the dose-length products. Shielding increased both the effective dose to the patient by 47% and the mean absorbed dose to the fetus (0.10 vs. 0.12 mGy; p < 0.001) compared with unshielded standard CTPA, as it affected the automatic exposure control. Shielded short CTPA marginally lowered only the mean fetal absorbed dose (0.03 vs. 0.02 mGy; p = 0.018). Shortening the scan reduced the fetal absorbed dose most effectively by 70% (0.10 vs. 0.03 mGy; p = 0.006), compared with the standard unshielded scan. Shielding modestly reduces fetal radiation dose but may compromise automatic exposure control, possibly increasing the maternal and fetal radiation dose. Shortening the scan is beneficial, assuming anatomical coverage is secured.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography