Solitaire Stentectomy Using a Stent-Retriever Technique in a Porcine Model
Andreas Simgen1, Michael Kettner2, Frida Juliane Webelsiep2
1Department of Neuroradiology, Saarland University Hospital, Kirrbergerstraße1, 66424, Homburg/Saar, Germany. andreassimgen@googlemail.com.
Clinical Neuroradiology
|May 20, 2020
Summary
Rescuing detached Solitaire devices during mechanical thrombectomy for ischemic stroke is feasible and effective. Different stent retrievers demonstrated comparable success rates and times without complications.
Area of Science:
- Endovascular therapy
- Neurointerventional radiology
- Stroke treatment
Background:
- Mechanical thrombectomy with Solitaire devices is standard for large vessel occlusion ischemic stroke.
- Inadvertent detachment of Solitaire devices is a feared complication with poor outcomes.
Purpose of the Study:
- To assess the feasibility and effectiveness of rescuing detached Solitaire devices.
- To evaluate different stent retrievers for Solitaire device retrieval in a porcine model.
Main Methods:
- Solitaire FR devices were deployed in the porcine axillary artery.
- Three different stent retrievers (Trevo ProVue, EmboTrap II, 3D) were used for rescue maneuvers.
- Rescue rates, time, and complications were assessed for 24 rescue attempts.
Main Results:
- 100% success rate for stentectomy (retrieval) of detached Solitaire devices.
- Comparable rescue times across all tested stent retrievers.
- No complications such as migration, vasospasm, or dissection were observed.
Conclusions:
- Stentectomy using various stent retrievers is a feasible and effective method for retrieving detached Solitaire devices.
- Trevo ProVue, EmboTrap II, and 3D revascularization devices showed similar efficacy and safety profiles.