Twiddler's syndrome with occult lead insulation damage in a child with epicardial pacemaker leads
Peter Kramer1, Felix Berger1,2,3, Björn Peters1
1Department of Congenital Heart Disease, Pediatric Cardiology, German Heart Centre Berlin, 13353Berlin, Germany.
Abstract:
We present a rare case of incidentally diagnosed Twiddler's syndrome in a child 7 years after implantation of a dual-chamber pacemaker system with epicardial leads. During revision, an insulation defect of the ventricular lead was evident, despite unremarkable prior pacemaker lead testing. The lead was repaired and a new generator was suture-fixated to prevent re-occurrence of generator manipulation.