Validation of Point of Care Hemoglobin Estimation Among Pregnant Women Using Digital Hemoglobinometers (HemoCue 301

Kapil Yadav1, Shashi Kant2, Gomathi Ramaswamy3

  • 14Room No: 33, Centre for Community Medicine, All India Institute of Medical Sciences, New Delhi, India.

Accurate and reliable point of care tests (POCTs) for hemoglobin estimation is essential for early diagnosis and management of anemia. This study was aimed to assess the reliability and validity of two versions of digital hemoglobinometer (HemoCue 201+ and HemoCue 301) compared to the gold standard Sysmex autoanalyzer for hemoglobin estimation. Pregnant women attending antenatal clinics of Primary Health Centre and a Sub Divisional Hospital in Haryana, India, were recruited. After obtaining consent, capillary blood samples were collected and tested for hemoglobin levels with digital hemoglobinometers (HemoCue 201+ and HemoCue 301). Among same pregnant women venous blood was collected and hemoglobin levels were estimated using autoanalyzer. Validity and reliability of POCTs compared to Sysmex autoanalyzer were reported. Of the 102 pregnant women included in the study, 44 (43%) were primigravida, with mean (SD) age of 23.3 (3.4) years. The mean (SD) of difference in hemoglobin levels using HemoCue 201+ was - 0.53 (1.01) and using HemoCue 301 was - 0.25 (0.85) g/dL as compared to auto-analyzer. Lin's concordance coefficient was 0.80 for HemoCue 201+ and 0.85 for HemoCue 301. Weighted Cohen's Kappa indicated moderate degree of agreement with the gold standard. Sensitivity (HemoCue 201+: 93%; HemoCue 301: 90%) and specificity (HemoCue 201: 76% HemoCue 301: 80%) for detecting anemia was similar for both the POCT devices. The digital hemoglobinometers used in the study had moderate degree of agreement and concordance with the autoanalyzer for hemoglobin estimation. HemoCue 301 had higher validity as compared to HemoCue 201+.

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.9K
Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.2K
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
988
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.3K