Characterization of a new scintillation imaging system for proton pencil beam dose rate measurements
Mahbubur Rahman1, Petr Brůža1, Katja M Langen2
1Thayer School of Engineering, Dartmouth College, Hanover, NH, United States of America.
Abstract:
The goal of this work was to create a technique that could measure all possible spatial and temporal delivery rates used in pencil-beam scanning (PBS) proton therapy. The proposed system used a fast scintillation screen for full-field imaging to resolve temporal and spatial patterns as it was delivered. A fast intensified CMOS camera used continuous mode with 10 ms temporal frame rate and 1 × 1 mm2 spatial resolution, imaging a scintillation screen during clinical proton PBS delivery. PBS plans with varying dose, dose rate, energy, field size, and spot-spacing were generated, delivered and imaged. The captured images were post processed to provide dose and dose rate values after background subtraction, perspective transformation, uniformity correction for the camera and the scintillation screen, and calibration into dose. The linearity in scintillation response with respect to varying dose rate, dose, and field size was within 2%. The quenching corrected response with varying energy was also within 2%. Large spatio-temporal variations in dose rate were observed, even for plans delivered with similar dose distributions. Dose and dose rate histograms and maximum dose rate maps were generated for quantitative evaluations. With the fastest PBS delivery on a clinical system, dose rates up to 26.0 Gy s-1 were resolved. The scintillation imaging technique was able to quantify proton PBS dose rate profiles with spot weight as low as 2 MU, with spot-spacing of 2.5 mm, having a 1 × 1 mm2 spatial resolution. These dose rate temporal profiles, spatial maps, and cumulative dose rate histograms provide useful metrics for the potential evaluation and optimization of dose rate in treatment plans.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016