Related Experiment Video
Updated: Dec 21, 2025

Author Spotlight: Developing a Point-of-Care Hemoglobin Estimation Method for Anemia Management
Published on: January 19, 2024
On-chip estimation of hematocrit level for diagnosing anemic conditions by Impedimetric techniques
Subhadip Chakraborty1, Sreyasi Das2, Chirantan Das1
1Department of Electronic Science, University of Calcutta, 92 A.P.C. Road, Kolkata, 700009, India.
Abstract:
An on-chip device has been fabricated on Si platform for precise estimation of the hematocrit (Hct) level of human blood with rapid turn out. Impedance/capacitance spectroscopy and current-voltage (I-V) measurements have been employed to observe the variation of electrical parameters of erythrocyte suspensions with varying Hct level. Experimentally obtained values of capacitance, impedance and conductance with Hct level suggests a linear variation. Current-time measurement has also been performed to ensure the susceptibility of red blood cells under a fixed external electric field for certain duration of time. The online real time sample analyzes have also been performed by the device connected with an embedded electronic circuit interfaced with a laptop through appropriate software. This has enabled the development of a novel Si-chip based compact point-of-care (POC) diagnosis system for Hct level estimation by measuring the dielectric/capacitive variation of an erythrocyte cell suspension. The relevant performance parameters of such a compact system including range, resolution, limit of detection and throughput have also been evaluated.