Objective:
This study aimed to optimize computed tomography (CT) parameters for detecting ground glass opacity nodules (GGNs) using a computer-assisted detection (CAD) system and a lung cancer screening phantom.
Methods:
A lung cancer screening phantom containing 15 artificial GGNs (-630 Hounsfield unit [HU], 2-10 mm) in the left lung was examined with a CT scanner. Three tube voltages of 80, 100, and 120 kVp were used in combination with five tube currents of 25, 50, 100, 200, and 400 mA; additionally, three slice thicknesses of 0.625, 1.25, and 2.5 mm and four reconstruction algorithms of adaptive statistical iterative reconstruction (ASIR-V) of 30, 60, and 90% were used. For each protocol, accuracy of the CAD system was evaluated for nine target GGNs of 6, 8, or 10 mm in size. The cut-off size was set to 5 mm to minimize false positives.
Results:
Among the 180 combinations of tube voltage, tube current, slice thickness, and reconstruction algorithms, combination of 80 kVp, 200 mA, and 1.25-mm slice thickness with an ASIR-V of 90% had the best performance in the detection of GGNs with six true positives and no false positives. Other combinations had fewer than five true positives. In particular, any combinations with a 0.625-mm slice thickness had 0 true positive and at least one false positive result.
Conclusion:
Low-voltage chest CT with a thin slice thickness and a high iterative reconstruction algorithm improve the detection rate of GGNs with a CAD system in a phantom model, and may have potential in lung cancer screening.
More Related Videos
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...