Perturbation analysis of 4D dose distribution for scanned carbon-ion beam radiotherapy
1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China; Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan; University of Chinese Academy of Sciences, Beijing 100049, China.
Purpose:
To evaluate the patients' set-up error-induced perturbation effects on 4D dose distributions (4DDD) of range-adapted internal target volume-based (raITV) treatment plan using lung and liver 4DCT data sets.
Methods:
We enrolled 20 patients with lung and liver cancer treated with respiratory-gated carbon-ion beam scanning therapy. PTVs were generated by adding a 2 mm range-adapted set-up margin on the raITVs. Set-up errors were simulated by shifting the beam isocenter in three translational directions of ±2 mm, ±4 mm, and ±6 mm. 4DDDs were calculated for both nominal and isocenter-shifted situations. Dose metrics of CTV dose coverage (D95) and normal tissue sparing were evaluated. Statistical significance with p < 0.01 was considered by Wilcoxon signed rank test.
Results:
The CTV dose coverage was more sensitive to set-up errors for lung cases than for liver cases, and more serious in superior-inferior direction. The sufficient CTV-D95 > 98% could be achieved with set-up errors less than ±2 mm in all shift directions both for lung and liver cases. With the increase of set-up error, the CTV dose coverage decreased gradually. The clinical criterial of CTV-D95 > 95% could not be fulfilled with set-up error reached to ±4 mm for lung cases, and ±6 mm for liver cases. OAR doses did not have a significant difference with each set-up error for both lung and liver cases.
Conclusions:
The range-adapted set-up margin successfully prevented dose degradation of 4DDDs in the presence of the same magnitude of set-up error for raITV-based carbon-ion beam scanning therapy.