Knee Wear Assessment: 3D Scanners Used as a Consolidated Procedure
Saverio Affatato1, Maria Cristina Valigi2, Silvia Logozzo2,3
1Laboratorio di Tecnologia Medica, IRCCS Istituto Ortopedico Rizzoli, Via di Barbiano, 1/10, 40136 Bologna, Italy.
Abstract:
It is well known that wear occurring in polyethylene menisci is a significant clinical problem. At this regard, wear tests on biomaterials medical devices are performed in order to assess their pre-clinical performance in terms of wear, durability, resistance to fatigue, etc. The objective of this study was to assess the wear of mobile total knee polyethylene inserts after an in vitro wear test. In particular, the wear behavior of mobile bearing polyethylene knee configurations was investigated using a knee joint wear simulator. After the completion of the wear test, the polyethylene mobile menisci were analyzed through a consolidated procedure by using 3D optical scanners, in order to evaluate the 3D wear distribution on the prosthesis surface, wear depths, wear rates, amount of material loss and contact areas. The results in terms of wear rates and wear volumes were compared with results of gravimetric tests, finding equivalent achievements.