Free-breathing BLADE acquisition method improves T2-weighted cardiac MR image quality compared with conventional
Jingjing Liu1,2, Hang Jin1, Yinyin Chen1
1Department of Radiology, Zhongshan Hospital, Fudan University and Shanghai Institute of Medical Imaging, Shanghai, PR China.
Background:
Cardiac magnetic resonance (MR) has become an essential diagnostic imaging modality in cardiovascular disease. However, the insufficient image quality of traditional breath-hold (BH) T2-weighted (T2W) imaging may compromise its diagnostic accuracy.
Purpose:
To assess the efficacy of the BLADE technique to reduce motion artifacts and improve the image quality.
Material And Methods:
Free-breathing TSE-T2W imaging sequence with cartesian and BLADE k-space trajectory were acquired in 20 patients. Thirty patients underwent conventional BH turbo spin-echo (TSE) T2W imaging and free-breathing BLADE T2W (FB BLADE-T2W) imaging. Twenty-one patients who had a signal loss of myocardium in BH short-axis T2W turbo inversion recovery (TSE-T2W-TIR) were scanned using free-breathing BLADE T2W turbo inversion recovery (BLADE TSE-T2W-TIR). The overall image quality, blood nulling, and visualization of the heart were scored on a 5-point Likert scale. The signal loss of myocardium, incomplete fat suppression near the myocardium, and the streaking or ghosting artifacts were noted in T2W-TIR sequences additionally.
Results:
The overall imaging quality, blood nulling, and the visualization of heart structure of FB BLADE-T2W imaging sequence were significantly better than those of FB T2W imaging with Cartesian k-space trajectory and BH TSE-T2W imaging sequence (P<0.01). The FB BLADE TSE-T2W-TIR reduces the myocardium signal dropout (P<0.05), incomplete fat suppression near myocardium (P<0.05), and the streaking and ghosting artifacts (P<0.05) in comparison with the BH TSE-T2W-TIR.
Conclusions:
FB BLADE T2W imaging provides improved myocardial visibility, less motion sensitivity, and better image quality. It may be applied in patients who have poor breath-holding capability.