Interference of Asfotase Alfa in Immunoassays Employing Alkaline Phosphatase Technology
Isabelle Danielle Piec1, Beatrice Tompkins1, William Duncan Fraser1
1University of East Anglia, Bob Champion Research and Education Building, Norwich, NR4 7UQ, United Kingdom.
Background:
Asfotase alfa (STRENSIQ®, Alexion Pharmaceuticals, Inc.) is the only approved treatment for patients with pediatric-onset hypophosphatasia, a disease caused by a mutation in the tissue-nonspecific alkaline phosphatase (TNSALP) gene. ALP is often used as signaling system in routine immunoassays. Because asfotase alfa contains the active site of the full ALP enzyme, it can catalyze the substrate as the antibody-conjugated ALP would within an assay. Therefore, its presence in a treated patient's sample may generate false positive or false negative results. We investigated whether the presence of asfotase alfa within a sample induced interference in immunoassays that utilize ALP or alternative detection systems.
Methods:
Asfotase alfa was added to samples at concentrations from 0.08-5 µg/mL and analysed on various immunoassays following manufacturer's instructions.
Results:
Asfotase alfa was detected in all ALP assays but ALKP1 (RayBiotech). We observed no changes in normetanephrine and noradrenaline (IBL) at any asfotase alfa concentration. However, asfotase alfa notably interfered in an oxytocin (ENZO) assay in nonextracted samples. Extraction using a C18 column eliminated the interference. No interference was observed on automated analyzers using alternative detection system (COBAS fT4 and TSH; Advia Centaur FSH, fT4; Architect LH; FSH). Immulite 2000 fT4, TSH, testosterone and hCG (ALP-based) showed no interference. However, the presence of asfotase alfa resulted in a dose-dependent increase of Troponin I signal.
Conclusion:
The presence of asfotase alfa must be taken into consideration when analyzing blood samples in treated patients to avoid any risk of misinterpretation of false positive/negative results. It is essential that assays be tested for this possible interference.
More Related Videos
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
10:05Simultaneous Distinction of Monospecific and Mixed DFS70 Patterns During ANA Screening with a Novel HEp-2 ELITE/DFS70 Knockout Substrate
Published on: January 17, 2018
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...