Method of Blood Pressure Measurement in Neonates and Infants: A Systematic Review and Analysis
Janis M Dionne1, Stephen A Bremner2, Simin K Baygani3
1Division of Nephrology, Department of Pediatrics, University of British Columbia, BC Children's Hospital, Vancouver, British Columbia, Canada.
Objectives:
To determine the recommended blood pressure (BP) measurement methods in neonates after systematically analyzing the literature regarding proper BP cuff size and measurement location and method.
Study Design:
A literature search was conducted in MEDLINE, PubMed, Embase, Cochrane Library, and CINAHL from 1946 to 2017 on BP in neonates <3 months of age (PROSPERO ID CRD42018092886). Study data were extracted and analyzed with separate analysis of Bland-Altman studies comparing measurement methods.
Results:
Of 3587 nonduplicate publications identified, 34 were appropriate for inclusion in the analysis. Four studies evaluating BP cuff size support a recommendation for a cuff width to arm circumference ratio of approximately 0.5. Studies investigating measurement location identified the upper arm as the most accurate and least variable location for oscillometric BP measurement. Analysis of studies using Bland-Altman methods for comparison of intra-arterial to oscillometric BP measurement show that the 2 methods correlate best for mean arterial pressure, whereas systolic BP by the oscillometric method tends to overestimate intra-arterial systolic BP. Compared with intra-arterial methods, systolic BP, diastolic BP, and mean arterial pressure by oscillometric methods are less accurate and precise, especially in neonates with a mean arterial pressure <30 mm Hg.
Conclusions:
Proper BP measurement is critical in neonates with naturally lower BP and attention to BP cuff size, location, and method of measurement are essential. With decreasing use of intra-arterial catheters for long-term BP monitoring in neonates, further studies are urgently needed to validate and develop oscillometric methodology with enhanced accuracy.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: