Influence of sample centrifugation on plasma platelet count and activated partial thromboplastin time using patient
Amalie Jesting1, Katja Kemp Jacobsen2, Aaron De Cock3
1University College Copenhagen, Department of Technology, Faculty of Health and Technology, Denmark; Department of Clinical Biochemistry, Rigshospitalet, Copenhagen University Hospital, Denmark.
Background:
Diagnostic coagulation testing is vulnerable to factors of the pre-analytical phase such as sample centrifugation. Despite this, centrifugation conditions differ widely among European laboratories. Here we use samples from patients referred for Activated partial thromboplastin time (APTT) testing to investigate if different centrifugation conditions result in platelet-poor plasma (PPP) (plasma platelet count < 10 × 109/L) and how the variation in centrifugation conditions affect APTT measurements.
Methods:
Centrifugation of 2000g (10 min) were compared with 3000g (10 min) using samples from patients referred for APTT testing (n = 70). Plasma platelet count and APTT were measured to investigate the influence of the centrifugation conditions. Differences were evaluated using Bland Altman Plots and Student's t-test.
Results:
Centrifugation at 3000g for 10 min produced PPP for more of the samples (64%) than centrifugation at 2000g (6%) (p < 0.001). No statistically significant difference for APTT (p = 0.265) was found for samples with APTT < 37 s while samples with prolonged APTT (>37 s) showed a statistically significant difference (p = 0.025). The Bland Altman plot did not reveal a clinically significant difference (mean difference 0.30 s/0.68%) when compared to a maximum acceptable bias of 10%.
Conclusion:
None of the centrifugation conditions used in this study adequately secured PPP for all samples. Despite a statistically significant difference between samples with prolonged APTT, no clinically significant difference was observed when comparing all APTT measurements.