Related Experiment Video
Updated: Dec 20, 2025

Recapitulation of an Ion Channel IV Curve Using Frequency Components
Published on: February 8, 2011
Accounting for variability in ion current recordings using a mathematical model of artefacts in voltage-clamp
Chon Lok Lei1, Michael Clerx1, Dominic G Whittaker2
1Computational Biology & Health Informatics, Department of Computer Science, University of Oxford, Oxford, UK.
Abstract:
Mathematical models of ion channels, which constitute indispensable components of action potential models, are commonly constructed by fitting to whole-cell patch-clamp data. In a previous study, we fitted cell-specific models to hERG1a (Kv11.1) recordings simultaneously measured using an automated high-throughput system, and studied cell-cell variability by inspecting the resulting model parameters. However, the origin of the observed variability was not identified. Here, we study the source of variability by constructing a model that describes not just ion current dynamics, but the entire voltage-clamp experiment. The experimental artefact components of the model include: series resistance, membrane and pipette capacitance, voltage offsets, imperfect compensations made by the amplifier for these phenomena, and leak current. In this model, variability in the observations can be explained by either cell properties, measurement artefacts, or both. Remarkably, by assuming that variability arises exclusively from measurement artefacts, it is possible to explain a larger amount of the observed variability than when assuming cell-specific ion current kinetics. This assumption also leads to a smaller number of model parameters. This result suggests that most of the observed variability in patch-clamp data measured under the same conditions is caused by experimental artefacts, and hence can be compensated for in post-processing by using our model for the patch-clamp experiment. This study has implications for the question of the extent to which cell-cell variability in ion channel kinetics exists, and opens up routes for better correction of artefacts in patch-clamp data. This article is part of the theme issue 'Uncertainty quantification in cardiac and cardiovascular modelling and simulation'.
Related Concept Videos
Patch Clamp
In this method, a glass micropipette containing electrolyte solution is tightly sealed against a small portion of the cell membrane. As a result, a patch of the cell...
Voltammograms: Overview
Shapes of Voltammograms
Voltammetry: Factors Affecting Measurements
Controlled-Current Coulometry: Overview